CYCLE-TAP Sampler Operating Instructions
Information
Question | How do I connect and operate the CYCLE-TAP Sampler? |
Answer |
The CYCLE-TAP sampler is a manually operated device for removing small volumes of solvent from the CYCLE-TAINER solvent delivery system using a syringe1. The sampler is designed to remove solvent aliquots directly from the container, completely eliminating contact with the atmosphere. The sampler features color-coded Quick-Connects allowing easy, air-tight attachment to the CYCLE-TAINER system and a three-way valve for purging the device to ensure the purity of the solvent sample is not compromised. The CYCLE-TAP sampler is available for use only with the CYCLE-TAINER solvent delivery system. The sampler is the easiest and safest way to obtain a small aliquot of solvent from the CYCLE-TAINER system, whether the requirement is for routine sampling or critical experimentation. By maintaining a closed system, solvent is not exposed to air where oxygen and water vapor can affect solvent quality. The sampler can be used to:
Using the CYCLE-TAP Sampler Attach the CYCLE-TAP sampler hose to the CYCLE-TAINER system solvent port using the color-coded Quick-Connect. Connect a gas source to the three-way ball valve on the sampler. Before sampling, purge the sampler with solvent and subsequently with gas. Connect a Luer lock syringe1 to the sampler and turn the three-way ball valve to the open position. Extract the syringe plunger to withdraw solvent from the container. Once the desired volume of solvent has been withdrawn, the three-way ball valve should be turned off and the syringe removed from the sampler. 1Syringe must be purchased separately. Avantor Performance Materials does not sell syringes. CYCLE-TAP Sampler Part # SolventQuick-Connect Description Recommended Luer Lock Syringe and Syringe Parts 25mL Syringe Pre-fitted with Luer Lock Valve Part Numbers: |
File | |
Keyword | CTS |