Information

Question What do I need in order to set up the 1250 L CYCLE-TOTE system?
Answer

There are several configuration options for CYCLE-TOTE system. At minimum, you will need:

  • Gas supply to blanket the solvent and avoid contact with air.
  • A stainless steel pump with a PTFE diaphragm to draw the solvent from the container.
  • Gas regulator to regulate the gas supply pressure.
  • Tubing and Quick-Connects to direct gas and solvent vent line for pressure relief valve vent line for pressure rupture disk.
  • Solvent line.
  • Gas line.
  • Ventilation may be needed during operation in the event of over pressurization of the container.
  • Grounding to eliminate static electricity build up near the flammable solvent.
  • An area for proper storage of the container. The CYCLE-TOTE system should be stored upright in a cool, clean, covered area.